STISP-800 Insulin Pimp Technical Specification | |
Item | Parameter |
Loading mode | Refill cartridge directly(no need reservoir) |
Size | 78mm×52mm×20mm |
Weight | 77g(Not included AAAbattery) |
Bluetooth | Support |
APP | Support,Check the usage of the insulin pump through the APP |
Device management | APP can customize the editing equipment name, department, bed number, patient name |
Blood sugar management | The APP can record the patient's blood sugar data, which is convenient for blood sugar monitoring |
Data sharing | Data can be shared with doctors or family members through the APP |
Data management | APP can generate treatment day report and cycle report, support wireless LAN one-key printing |
Data download | Support |
PC-side management | Support, you can check the usage of the insulin pump through the computer |
PC-side operation | The insulin pump treatment plan can be set on the computer, including: basal rate, high-dose preset value for three meals, pulse cycle, daily total limit, high-dose limit, alarm mode, meal reminder, blood sugar measurement reminder, etc. |
CPU | Dual CPU |
Minimum dose | 0.05U |
Waterproof | IPX7 |
Battery | 1 AAA Alkaline battery+Built-in Lithium battery |
Menu | Open |
Infusion method | Basal infusion, clinical-based infusion, high-dose infusion, square wave infusion |
Charging method | The screw is automatically reset, and the loaded amount of medicine is automatically calculated. |
Base rate segment | 24/48 basal rates |
Base rate setting step amount | 0.1U/0.05U |
APP setting | Support |
Quick setting of base rate | 24-segment and 6-segment dual database |
Custom base rate database | APP setting |
Base rate plan | 3 sets can be stored in advance |
Base rate calibration function | Yes |
Base rate calibration function | Yes |
High-dose setting mode | Regular mode and three-meal mode |
High-dose preset function | Yes |
High Dose Wizard Calculator | Yes |
Blood sugar target | Up to 48 |
Bolus infusion rate | Adjustable, 4 pulse periods:5s,10s,15s,20s |
Large dose setting increments | 0.1U(0-10U),0.5U(10-99U) |
High dose limit | 20u (default),Setting 0-88U |
Daily total limit | 60u (default),Setting 0-200U |
Daily total review record | 500 PCS record |
Basal rate review record | 500 PCS record |
High dose review record | 500 PCS record |
Exhaust review record | 500 PCS record |
alarm Review record | 500 PCS record |
Warning item | Obstruction, Low dose, drug out, low battery, Meal, Blood sugar measurement, Alarm |
Alarm method | Buzzer and vibration alarm |
Meal Tips | Can be opened, meal reminder 10min/20min/30min after high-dose infusion |
Blood sugar test tips | It can be turned on, and the blood sugar measurement prompt is 1h/2h/3h after high-dose infusion |
Maximum charge | 305U |
Reservoir volume | 3.0mL |
Warning item parameters: | |
Low voltage | Low voltage prompt when the voltage is lower than 1.2V |
Low dose | When the liquid medicine is lower than 20U, the low liquid medicine prompt will be displayed. |
Drug out | When there is no infusion of liquid medicine, it will alarm when the medicine is finished |
Block | Blockage alarm before infusion line blockage reaches 6U |
Certificates
CE,ISO,Free sales certificate,Registeration certificate for medical device,Manufacturing license,qualified industrial registerations & ect .Packing
Transportation